A user facility biomedical technician (biomed) reported finding burnt cables on an aquabplus 2000.
Upon follow-up, it was confirmed that thermal damage was found on the cables between the motor protection switch and the power contactor.
The burnt cables were identified during the technical safety check.
To resolve the reported issue, the biomed replaced all three cables.
However, the following day the biomed realized the pump was rotating backwards.
When reassembling the parts (post repair), were placed in the wrong positions.
To fix this, the biomed swapped them and the pump began rotating in the correct direction.
At the time of follow-up, the system was reported to be fully operational.
The biomed stated there have been no further issues.
The damaged cables were reported to be available for evaluation.
The biomed was unable to provide ftp machine files review because the sd card was reportedly corrupt.
There was no patient involvement associated with the reported event.
|