• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC EUROPE SARL SERENA QUAD CRT-P MRI SURESCAN PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC EUROPE SARL SERENA QUAD CRT-P MRI SURESCAN PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number W4TR02
Device Problems False Alarm (1013); Low impedance (2285)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/01/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that there was undersensing on the right atrial (ra) lead. It was also noted there was an alert for right atrial (ra) lead low impedance on the device, however, when interrogated the lead trends showed no out of range impedances. The lead and device remain in use. No patient complications have been reported as a result of this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSERENA QUAD CRT-P MRI SURESCAN
Type of DevicePULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ 1131
Manufacturer (Section G)
MEDTRONIC EUROPE SARL
route du molliau 31
case postale
tolochenaz vaud 1131
SZ 1131
Manufacturer Contact
paula bixby
8200 coral sea st ne
mounds view, MN 55112
7635055378
MDR Report Key11869929
MDR Text Key252157535
Report Number9614453-2021-02122
Device Sequence Number1
Product Code NKE
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P010015
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 08/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date12/28/2021
Device Model NumberW4TR02
Device Catalogue NumberW4TR02
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received07/15/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/07/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 05/24/2021 Patient Sequence Number: 1
Treatment
439888 LEAD, 5076-45 LEAD, 5076-52 LEAD.
-
-