Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Post Operative Wound Infection (2446); Limb Fracture (4518)
Event Date 03/10/2021
Event Type  Injury  
Manufacturer Narrative
A product investigation was performed for this device.The actual device was not returned to the manufacturer for evaluation.The root cause of the non-union is undetermined.There is no way to predict a non-union or failure to heal.The root cause of the infection is undetermined.Infections have many causes and treatments.Each infection is addressed individually and taken very seriously by both the attending surgeon and sign surgeons at sign headquarters.The radiographic and clinical data were reviewed by a sign orthopedic surgeon.This failure does not indicate a defect in the product.The risk associated was evaluated during the risk management process and was deemed an acceptable level of risk to the patient.Sign fracture care international continues to monitor these events as part of our post market surveillance activities.
 
Event Description
We became aware on 5/13/2021 that a sign im nail implanted to repair a fracture was replaced due to a non-union with infection.The im nail was replaced with a 9mm x 400mm standard im nail per the sign technique manual.Surgeon comment: "re-admission for dbr and exchange nail due to sis infection.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SIGN IM NAIL
Type of Device
INTRAMEDULLARY FIXATION ROD
Manufacturer (Section D)
SIGN FRACTURE CARE INTERNATIONAL
451 hills street
suite b
richland WA 99354 5502
Manufacturer (Section G)
SIGN FRACTURE CARE INTERNATIONAL
451 hills street
suite b
richland WA 99354 5502
Manufacturer Contact
robert schmitt
451 hills street
suite b
richland, WA 99354-5502
5093711107
MDR Report Key11871379
MDR Text Key252210878
Report Number3034525-2021-00086
Device Sequence Number1
Product Code HSB
Combination Product (y/n)N
Reporter Country CodeGA
PMA/PMN Number
K022632
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Reporter Occupation Physician
Type of Report Initial
Report Date 05/13/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/24/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator No Information
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received05/13/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age24 YR
Patient Weight91
-
-