OLYMPUS MEDICAL SYSTEMS CORP. FIBERSCOPE "CYF-5", EUROPEAN VERSION; CYSTONEPHROFIBERSCOPE Back to Search Results
Model Number CYF-5
Device Problems Loose or Intermittent Connection (1371); Poor Quality Image (1408)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device was returned, and an initial evaluation was conducted by olympus; however, investigation is ongoing.During the initial evaluation, the user's report was successfully duplicated.Additionally, there were cracks noted on the distal end glue and a small channel leak noted during the scope leak test.If additional information becomes available following device evaluation, a supplemental report will be filed.
 
Event Description
It was reported, the biopsy channel on the olympus fiberscope was loose and producing a cloudy image.There was no patient harm or consequence reported as a result of this event.
 
Manufacturer Narrative
This report is being supplemented to provide additional information based on the legal manufacturer's final investigation.A review of the device history record (dhr) found no deviations that could have caused or contributed to the reported issue.Based on the results of the investigation, the probable cause of the biopsy channel loose causing a cloudy image is likely due to handling.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FIBERSCOPE "CYF-5", EUROPEAN VERSION
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11885124
MDR Text Key268350958
Report Number8010047-2021-06750
Device Sequence Number1
Product Code FAJ
UDI-Device Identifier04953170339417
UDI-Public04953170339417
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 07/02/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/07/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Date Manufacturer Received06/25/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured11/02/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-