MEDTRONIC PUERTO RICO OPERATIONS CO. INTELLIS STIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR Back to Search Results
Model Number 97715
Device Problems Failure to Interrogate (1332); Overheating of Device (1437); Communication or Transmission Problem (2896); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Burn(s) (1757); Burning Sensation (2146); No Clinical Signs, Symptoms or Conditions (4582)
Event Type  Injury  
Manufacturer Narrative
Concomitant medical products: product id 97755 lot# serial# unknown implanted: explanted: product type recharger. Information references the main component of the system. Other relevant device(s) are: product id: 97755, serial/lot #: unknown, ubd: , udi#: if information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a patient (pt) who was implanted with an implantable neurostimulator (ins) for unknown indications for use. It was reported that their recharger has been becoming hot while charging their implant. Pt stated that they're blind so they don't know if their skin was actually burnt but pt said it feels like a sunburn on their skin while they're charging and after they remove the paddle from their skin. Pt stated they tried to put clothing in between the recharger and their skin but they couldn't get a connection unless they put it directly on their skin. Pt stated this started happening about 4-5 months ago.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameINTELLIS
Type of DeviceSTIMULATOR, SPINAL-CORD, TOTALLY IMPLANTED FOR
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
david gustafson
7000 central avenue ne rcw215
minneapolis, MN 55432
7635149628
MDR Report Key11886852
MDR Text Key267300532
Report Number3004209178-2021-08393
Device Sequence Number1
Product Code LGW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P840001
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer
Reporter Occupation
Type of Report Initial,Followup,Followup
Report Date 09/17/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/26/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date06/14/2019
Device Model Number97715
Device Catalogue Number97715
Was Device Available for Evaluation? No
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/16/2021
Was Device Evaluated by Manufacturer? No Answer Provided
Date Device Manufactured06/15/2018
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 05/26/2021 Patient Sequence Number: 1
-
-