MEDIVANCE, INC. ¿ 1725056 ARCTIC SUN® 5000 ARCTIC SUN DEVICE Back to Search Results
Model Number 50000000
Device Problems Insufficient Cooling (1130); Insufficient Information (3190)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 05/04/2021
Event Type  Injury  
Manufacturer Narrative
The investigation is still in progress. Once the investigation is complete a supplemental report will be filed.
 
Event Description
Hold for review tc 5. 27. 21. It was reported that patient (weighed (b)(6)) was actively shivering and above target. Nurse asked for recommendations. Target was 36 c in hypothermia mode. Patient temperature was 37 c, wt was 4. 5 c, flow rate was 2. 5 l/min. Therapy was initiated at 1 am. Mag and fentanyl drugs was given. Counter warming measures were in place. There were 5 pads in place, only one universal. Doctor discussed with rn, the reasons for heat generation and recommended adding another universal pad and reaching out to medical team to address active shivering. Patient was on paralytic and 2nd universal pad was added. Pt was 36. 4 c and wt was 11. 1 c.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARCTIC SUN® 5000
Type of DeviceARCTIC SUN DEVICE
Manufacturer (Section D)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer (Section G)
MEDIVANCE, INC. ¿ 1725056
321 s taylor ave
suite 200
louisville 80027
Manufacturer Contact
yonic anderson
8195 industrial blvd
covington 30014
7707846100
MDR Report Key11895686
MDR Text Key253049926
Report Number1018233-2021-03071
Device Sequence Number1
Product Code DWJ
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K161602
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Health Professional,User Facility
Reporter Occupation
Type of Report Initial,Followup
Report Date 11/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/27/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator
Device Model Number50000000
Device Catalogue Number50000000
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/19/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/01/2015
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 05/27/2021 Patient Sequence Number: 1
-
-