Model Number CYF-5
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject was returned to the service department of (b)(4) but has not returned to omsc.The service department of (b)(4) checked the subject device for the evaluation.It was confirmed that the wire came out from the bending section rubber which seemed to be occurred due to the user handling or wear and tear.Omsc reviewed the manufacturing history (dhr) of the subject device and confirmed no irregularity.The exact cause of the reported event could not be conclusively determined.However, based on the report of the service department of (b)(4) and a thing which the bending section rubber of the subject device was torn, omsc surmised there was the possibility this phenomenon was attributed to applying the external force to the distal bending section rubber.In addition, we have confirmed that this reported phenomenon was not caused by the device or the manufacture, and there were no problems with the safety and there is no frequent occurrence.If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp.(omsc) was informed that during the incoming inspection for the repair at the service department of olympus (b)(4), it was found the wire came out from the bending section rubber of the subject device.There was no patient injury associated with this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OES CYSTONEPHROFIBERSCOPE
Type of Device
CYSTONEPHROFIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11900370
MDR Text Key253150955
Report Number8010047-2021-06850
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeGM
PMA/PMN Number
K032092
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 05/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberCYF-5
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received05/04/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/12/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-