• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW MEDICAL LTD. IV3000 6X7 CARTON OF 100; TAPE AND BANDAGE, ADHESIVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW MEDICAL LTD. IV3000 6X7 CARTON OF 100; TAPE AND BANDAGE, ADHESIVE Back to Search Results
Model Number 66024007
Device Problem Peeled/Delaminated (1454)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/17/2021
Event Type  malfunction  
Event Description
It was reported that, during treatment, nineteen (19) of iv3000 6x7 peeled off, are wrinkled and cannot be used.Treatment was resumed, without any significant delay, with an available smith and nephew back up device; therefore, patient was not harmed.
 
Manufacturer Narrative
H3, h6: the batch record was reviewed and it could be confirmed that the products had progressed through a satisfactory release process which involved testing the product against the requirements of the finished product specification.There was also no issue identified during the manufacturing process that could have caused or contributed to the complaint problem.A complaint history review found a small number of similar instances of the reported event.There is nothing to indicate that this is outside of acceptable rates of occurrence.The device was used for treatment and was not returned for analysis.We have not been able to confirm a relationship between the event and the device or identify a definitive root cause.It was reported that the dressing peeled off during treatment.Probable root causes include environmental issues such as temperature, as temperature can alter the adhesive nature of the dressing and also the incorrect skin preparation and application of dressings.Dressing should be stored in the conditions detailed in the ifu.The skin should also be prepared and dressings applied in accordance with instructions detailed in the ifu.The users of the reported product are therefore advised to consult the ifu, to delineate future occurrences of the reported issue.This guide provides comprehensive instructions of the operation, use and limitations of the device, including application and removal of dressings, skin preparation and conditions in which the device should be stored.No further actions by smith and nephew are deemed necessary at this stage.However, we will continue to monitor for any adverse trends relating to this product range.Smith and nephew acknowledge customer concern and are continually investigating ways to develop and improve our products.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
IV3000 6X7 CARTON OF 100
Type of Device
TAPE AND BANDAGE, ADHESIVE
Manufacturer (Section D)
SMITH & NEPHEW MEDICAL LTD.
101 hessle road
hull east riding of yorkshire HU3 2 BN
UK  HU3 2BN
MDR Report Key11910642
MDR Text Key253753790
Report Number8043484-2021-01292
Device Sequence Number1
Product Code KGX
UDI-Device Identifier05000223417383
UDI-Public05000223417383
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Type of Report Initial,Followup
Report Date 10/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/31/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/01/2023
Device Model Number66024007
Device Catalogue Number66594007
Device Lot Number2013
Was Device Available for Evaluation? Yes
Date Manufacturer Received10/01/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-