Model Number 72200568
Device Problem Overheating of Device (1437)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/13/2021
Event Type  malfunction  
Manufacturer Narrative
Internal complaint reference (b)(4). The reported device, intended for use in treatment, was received for evaluation. A visual inspection was performed on the exterior of product and no physical damage was observed. There was a relationship found between the returned device and the reported incident. Complaint of e12 temp error was confirmed. Cause of e12 error is a defective power supply. Unit powered up and passed functional testing with a known good power supply installed. The complaint investigation has concluded the cause of the failure to be a defective electronic component. A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended. No containment or corrective actions are recommended at this time.
 
Event Description
It was reported that the light source had an e12 error code. No case involved; therefore, there was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameLIGHT SOURCE 500XL XENON
Type of DeviceIMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer (Section G)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key11910724
MDR Text Key254064066
Report Number1643264-2021-02028
Device Sequence Number1
Product Code FFS
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K994084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial
Report Date 05/31/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received05/31/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number72200568
Device Catalogue Number72200568
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/27/2021
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/17/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unkown

-
-