• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL Back to Search Results
Model Number 0998-00-0800-53
Device Problem Unexpected Shutdown (4019)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/11/2021
Event Type  malfunction  
Manufacturer Narrative
A getinge field service engineer (fse) was dispatched to evaluate this unit.The fse was unable to duplicate the issue.It was determined the patient was unintentionally pinching the catheter extension.The fse completed all safety, functionality and calibration checks and all tests passed to factory specifications.The iabp unit was cleared for clinical use and released to the customer.
 
Event Description
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) was pumping and intermittently stopped.The iabp was swapped for another one.No patient harm, serious injury or adverse event was reported.
 
Manufacturer Narrative
Updated fields: b4, g4, g7, h2, h6 (type of investigation), h10, h11.Corrected field: h4.Analysis of production: (3331/213) the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the dhr/lhr review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis: (4109/213) the review of the historical data indicates that no other similar complaint was reported for the same lot/serial number and reported failure mode.Trend analysis: (4110/213) the overall 24 month product complaint trend data for the period (jun 2019 through may 2021) was reviewed.There were no triggers identified for the review period.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CARDIOSAVE HYBRID TYPE B PLUG
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
MDR Report Key11931037
MDR Text Key254080247
Report Number2249723-2021-01171
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567108391
UDI-Public10607567108391
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 07/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/30/2021
Patient Sequence Number1
-
-