• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. REDUCER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. REDUCER; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71631124
Device Problem Corroded (1131)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/24/2021
Event Type  malfunction  
Event Description
It was reported that during set up for an internal fixation surgery, the reducers were noticed to have rust on them.There was no significant delay and the procedure was finished using a smith & nephew back up.Patient was not harmed.
 
Manufacturer Narrative
H6: the device, imtended for use in treatment, was not returned for evaluation.Therefore, the product analysis and the reported event could not be confirmed at this time.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.Should additional information be received, the complaint will be reopened.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
REDUCER
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11935451
MDR Text Key254257433
Report Number1020279-2021-04984
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010419606
UDI-Public03596010419606
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Type of Report Initial,Followup
Report Date 07/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71631124
Device Catalogue Number71631124
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-