Catalog Number M003112FPP0
Device Problem Premature Activation (1484)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/08/2021
Event Type  malfunction  
Event Description
It was reported that during the mca and ica aneurysm case a guide catheter was used to deploy the subject stent. Physician felt some resistance advancing the subject stent delivery system through the patient's anatomy. For deployment the inner shaft was pushed to the distal but there was some resistance and subject stent did not deployed. The physician tried to adjust the subject stent for several times but failed to do so and decided to remove it. While doing so the subject stent deployed prematurely inside the guide catheter. Patient's anatomy was severely tortuous. The physician decided that the subject device was not suitable for the patient and terminated the operation. No further information is available.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSURPASS STREAMLINE 4.0MM X 30MM - CE
Type of DeviceINTRACRANIAL ANEURYSM FLOW DIVERTER
Manufacturer (Section D)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI NA
Manufacturer (Section G)
STRYKER NEUROVASCULAR CORK
ida industrial estate
model farm road
cork NA
EI NA
Manufacturer Contact
tara lopez
47900 bayside parkway
fremont, CA 94538
5104132500
MDR Report Key11935740
MDR Text Key254486605
Report Number3008881809-2021-00222
Device Sequence Number1
Product Code OUT
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
P170024
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Reporter Occupation
Type of Report Initial,Followup
Report Date 10/14/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/03/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Catalogue NumberM003112FPP0
Device Lot Number22397034
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/20/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/22/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured08/17/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

-
-