(b)(4).Customer has indicated that the product will not be returned to zimmer biomet for the investigation as the product remains implanted.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034 - 2021 - 01716, 0001825034 - 2021 - 01717.
|
(b)(4).This follow-up report is being submitted to relay additional information.Reported event was confirmed by review of medial records.Device history record (dhr) review was unable to be performed as the lot number of the device involved in the event is unknown.Root cause was unable to be determined as the necessary information to adequately investigate the reported event was not provided.Medical records identified the following: metallosis/trunnionosis and pseudotumor formation.Patient presented with hip pain, swelling, and abductor dysfunction, infection workup negative with normal esr and crp.Pseudotumor diagnosed on mri, ct showed significant osteolysis around the cup but no obvious loosening.Pseudotumor was excised and all necrotic tissue was removed.Noted was near complete destruction of the gluteus medius and minimus with no attachment on the greater trochanter.Head removed from the stem without difficulty, trunnionosis with black corrosion on the femoral head taper as well as the stem, stem well fixed.Significant osteolysis along the superior and lateral acetabulum, no evidence of loosening.Cup left in place with bone graft packing.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|