ABBOTT VASCULAR ARMADA 35 PTA CATHETER PERIPHERAL DILATATION CATHETER Back to Search Results
Model Number B1080-060
Device Problems Off-Label Use (1494); Difficult to Remove (1528); Material Rupture (1546); Material Separation (1562); Improper or Incorrect Procedure or Method (2017)
Patient Problem Insufficient Information (4580)
Event Date 05/14/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4). The device was received. Investigation is not yet complete. A follow-up report will be submitted with all additional relevant information.
 
Event Description
It was reported that the procedure was to treat a re-stenosed lesion located in the right iliac vein that was non-calcified and non-tortuous. After the armada balloon dilatation catheter (bdc) was advanced to the lesion, and during the second inflation below nominal pressure, the armada balloon ruptured circumferentially. During removal of the device, there was resistance and device interaction within the stent; the balloon folded over on itself and the distal part of the balloon separated in the patient. A snare was used to successfully retrieve the separated balloon portion. A non-abbott balloon was used to complete the procedure. No additional information was provided. There was no adverse patient sequela and no clinically significant delay reported. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameARMADA 35 PTA CATHETER
Type of DevicePERIPHERAL DILATATION CATHETER
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR COSTA RICA, REG # 3009564766
52 calle 3 b31 coyol free zone
el coyol alajuela
CS
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key11947493
MDR Text Key254809021
Report Number2024168-2021-04706
Device Sequence Number1
Product Code LIT
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111899
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 06/23/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberB1080-060
Device Catalogue NumberB1080-060
Device Lot Number10217G1
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/25/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/17/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 06/07/2021 Patient Sequence Number: 1
-
-