Model Number N/A
Device Problem Mechanical Jam (2983)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/17/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Multiple mdr reports were filed for this event, please see associated report: 0001822565-2021-01574.Concomitant medical product: degree distal cut block, item# 42509901000, lot# 63842130.The product has been returned to zimmer biomet for investigation.The investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that a trocar pin got stuck in the cutting block.The block was removed from the patient and a new cutting block was used.There is no additional information available.
 
Event Description
No further event information available at the time of this report.
 
Manufacturer Narrative
This follow-up is being submitted to relay additional information.Visual examination of the returned products identified signs of repeated use for both devices.The trocar drill pin is seized in one of the guide holes of the distal cut guide.Dimensional analysis of the trocar drill pin determined that the product, where measured, was conforming to print specifications.Medical records were not provided.Device history record (dhr) was reviewed for deviations and/ or anomalies with no deviations / anomalies identified.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEADLESS TROCAR DRILL PIN 3.2 MM DIAMETER 75 MM LENGTH
Type of Device
ORTHOPEDIC STEREOTAXIC INSTRUMENT, KNEE
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11948073
MDR Text Key254636770
Report Number0001822565-2021-01575
Device Sequence Number1
Product Code OLO
UDI-Device Identifier00889024201620
UDI-Public(01)00889024201620(17)301212(10)64957279
Combination Product (y/n)N
PMA/PMN Number
K182964
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 07/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/07/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number00590102000
Device Lot Number64957279
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/26/2021
Was the Report Sent to FDA? No
Date Manufacturer Received07/20/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
Patient Weight80
-
-