Model Number 0998-00-0800-53
Device Problem Failure to Charge (1085)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/14/2021
Event Type  malfunction  
Manufacturer Narrative
At this time, the customer has not requested for getinge to evaluate the iabp unit involved in this event. A getinge stm has conversed with the customer by phone and stated that the pump console was not properly latched into the hospital cart. The on-site biomedical technician had reseated the pump into the cart and verified that the batteries were now charging properly. The request for a service call was cancelled. A supplemental report will be submitted upon completion of our investigation.
 
Event Description
It was reported that the batteries on the cardiosave intra-aortic balloon pump (iabp) were not charging when the unit was connected to the ac power. There was no patient involvement, and no adverse event reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameCARDIOSAVE HYBRID TYPE B PLUG
Type of DeviceSYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer (Section G)
DATASCOPE CORP. - MAHWAH
1300 macarthur blvd
mahwah NJ
Manufacturer Contact
dorota wolpiuk
1300 macarthur blvd
mahwah, NJ 
MDR Report Key11957340
MDR Text Key254897824
Report Number2249723-2021-01211
Device Sequence Number1
Product Code DSP
Combination Product (y/n)N
PMA/PMN Number
K112372
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 06/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/08/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number0998-00-0800-53
Device Catalogue Number0998-00-0800-53
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received05/17/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured05/15/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device?
Type of Device Usage Reuse

Patient Treatment Data
Date Received: 06/08/2021 Patient Sequence Number: 1
Treatment
N/A.
-
-