DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Failure to Charge (1085)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/14/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
At this time, the customer has not requested for getinge to evaluate the iabp unit involved in this event.
A getinge stm has conversed with the customer by phone and stated that the pump console was not properly latched into the hospital cart.
The on-site biomedical technician had reseated the pump into the cart and verified that the batteries were now charging properly.
The request for a service call was cancelled.
A supplemental report will be submitted upon completion of our investigation.
|
|
Event Description
|
It was reported that the batteries on the cardiosave intra-aortic balloon pump (iabp) were not charging when the unit was connected to the ac power.
There was no patient involvement, and no adverse event reported.
|
|