The reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A visual inspection revealed that both meniscal suture loops were detached from their handles.The complaint was confirmed, and the root cause was associated with device design.Factors that could have contributed to the reported event, include an application of unintended inappropriate or excessive force to the device or packaging, attempted correction of a damaged device, or an impact event inconsistent with normal use.A corrective action for this failure mode is in place.
|