• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: EDWARDS LIFESCIENCES EDWARDS ESHEATH INTRODUCER SHEATH; PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

EDWARDS LIFESCIENCES EDWARDS ESHEATH INTRODUCER SHEATH; PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED Back to Search Results
Model Number ESHEATH UNKNOWN
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hematoma (1884)
Event Date 12/02/2020
Event Type  Injury  
Manufacturer Narrative
This report provides data from thv/tvt registry exemption number e2016006 and summarizes 1 hematoma at access site serious injury event for the sapien 3 transcatheter heart valve in the mitral position.The 'time to event' (tte, in days) for this event was 0.00.The device identification (di) numbers for edwards esheath introducer set are (b)(4).The instructions for use (ifu) list hemorrhage requiring transfusion or intervention as a potential risk associated with the overall tavr procedure, including potential access complications associated with standard cardiac catheterization.The majority of transfemoral access related bleeding complications are related to diseased ileo-femoral vessels and/or procedural technique during the insertion or removal of the sheath and/or dilators.There are cases where the bleeding is significant and more complex intervention or transfusion is required to treat/prevent a permanent injury from occurring.The thv physician training manuals instruct on procedural considerations for sheath insertion with regards to proper screening critical to reducing vascular complications.The ifu contraindicates patients with ilio-femoral vessel characteristics that would preclude safe placement of sheaths such as severe obstructive calcification or severe tortuosity.Pre-procedure screening and assessment of the femoral/iliac artery internal diameters will enable the clinician to determine if the valve can be delivered transfemorally.In this case, specific procedural details are not available to determine potential contributing factors to the event, or if the event is related to an edwards device; however, a transfusion of 2 units or more of blood was reported in the registry.By varc definition, the meet criteria for major vascular complication; therefore, the event is being reported.The ifu and training manuals have been reviewed and no inadequacies have been identified with regards to warnings, contraindications, and the directions/conditions for the successful use of the device.Complaint histories for all reported events are reviewed against trending control limits on a monthly basis, and any excursions above the control limits are assessed and documented as part of this monthly review.No corrective or preventative actions are required at this time.
 
Event Description
Thv/tvt registry summary reporting for adverse event submission for quarter 1 2021 data extract for mitral serious injuries for the sapien 3 valve.This report summarizes 1 hematoma at access site serious injury event for the sapien 3 transcatheter heart valve.The age range for these events is (b)(6) years.The breakdown for gender is as follows: 1 female.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EDWARDS ESHEATH INTRODUCER SHEATH
Type of Device
PROSTHESIS, MITRAL VALVE, PERCUTANEOUSLY DELIVERED
Manufacturer (Section D)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer (Section G)
EDWARDS LIFESCIENCES
1 edwards way
irvine CA 92614
Manufacturer Contact
renee van dorne
1 edwards way
irvine, CA 92614
9492506385
MDR Report Key11977494
MDR Text Key255486343
Report Number2015691-2021-03456
Device Sequence Number1
Product Code NPU
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140031
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 06/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberESHEATH UNKNOWN
Device Catalogue NumberN/A
Device Lot NumberN/A
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received06/03/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
Patient Age72 YR
-
-