• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ACUMED LLC ACU-LOC VDU PLATE LEFT STD; PLATE, FIXATION, BONE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ACUMED LLC ACU-LOC VDU PLATE LEFT STD; PLATE, FIXATION, BONE Back to Search Results
Model Number 70-0045-S
Device Problems Entrapment of Device (1212); Physical Resistance/Sticking (4012)
Patient Problem Insufficient Information (4580)
Event Date 05/14/2021
Event Type  malfunction  
Manufacturer Narrative
The returned 2.0mm locking drill guide 4mm-32mm, 2.0mm quick release drill and acu-loc® vdu plate left std were examined visually under magnification.The drill guide cannot be manually unscrewed from the plate, and the drill bit is stuck with the cutting portion protruding from the bottom of the plate and the shaft sheared off at the end of the drill guide.The fracture pattern indicates a blunt and brittle fracture with jagged edges and some torsional markings at the center of the drill cross-section.Additionally, 360-degree markings can be seen at the tip of the head portion of the broken drill bit.The main thing to note is that the surgical technique instructs that the radiopaque targeting guide and locking bolt be used with the acu-loc plates to ensure proper placement and trajectory of the drill guide, drill, and screws.Improper trajectory of the drill guide into the plate can cause off-axis cross-threading, leading to the drill guide cutting into the plate and getting stuck.If the drill gets stuck within dense material as well as within an improperly assembled plate and drill guide, the combined forces can cause the drill bit to break when turning on the drill.Additional mdrs associated with this event: 3025141-2021-00049: drill, 3025141-2021-00051: screw, 3025141-2021-00052: drill guide.
 
Event Description
While implanting a aculoc vdu plate in the patient's wrist, the drill became stuck in the drill guide and could not be pulled out.The whole plate was removed and a second plate was successfully implanted.It is likely the drill was at an angle in the drill guide making it get stuck.There was a 15 minute delay in surgery.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACU-LOC VDU PLATE LEFT STD
Type of Device
PLATE, FIXATION, BONE
Manufacturer (Section D)
ACUMED LLC
5885 ne cornelius pass road
hillsboro OR 97124
Manufacturer (Section G)
ACUMED LLC
5885 ne cornelius pass road
hillsboro OR 97124
Manufacturer Contact
micki anderson
5885 ne cornelius pass road
hillsboro, OR 97124
8886279957
MDR Report Key11980036
MDR Text Key260555147
Report Number3025141-2021-00053
Device Sequence Number1
Product Code HRS
Combination Product (y/n)N
Reporter Country CodeUK
PMA/PMN Number
K102998
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 05/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/10/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/26/2024
Device Model Number70-0045-S
Device Catalogue Number70-0045-S
Device Lot Number403772
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/08/2021
Date Manufacturer Received05/14/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured02/06/2017
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-