OLYMPUS MEDICAL SYSTEMS CORP. HALOGEN LIGHT SOURCE Back to Search Results
Model Number CLK-4
Device Problems Electrical /Electronic Property Problem (1198); Failure to Power Up (1476)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The subject was returned to the service department of oekg but has not returned to omsc. The service department of oekg checked the subject device for evaluation. Oekg found the reported phenomena. Also, it was found that there were wear and tear defects, and the cracks on front panel. The exact cause of the reported event could not be conclusively determined at this time. If additional information is received, this report will be supplemented.
 
Event Description
Olympus medical systems corp. (omsc) was informed that during the incoming inspection for the repair at the service department of olympus (b)(6) (oekg), the subject device could not be switched on because the power switch did not work. Also it was found that the lamp of the subject device was failed, and the spare lamp was missing. There was no patient injury associated with this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHALOGEN LIGHT SOURCE
Type of DeviceHALOGEN LIGHT SOURCE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA 192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key11989249
MDR Text Key270491007
Report Number8010047-2021-07450
Device Sequence Number1
Product Code FAJ
Combination Product (y/n)N
Reporter Country CodeBE
PMA/PMN Number
K993041
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation
Type of Report Initial,Followup
Report Date 07/30/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/13/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model NumberCLK-4
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/28/2021
Is the Reporter a Health Professional? No
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/07/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/04/2013
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No