Catalog Number 106697 |
Device Problems
Crack (1135); Leak/Splash (1354)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/20/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(6).Should additional relevant information become available, a supplemental report will be submitted.
|
|
Event Description
|
It was reported that during that an external fluid leak was observed from a small crack in the lower part of a prismaflex m100 set.No alarm was generated.There was no patient involvement.No additional information is available.
|
|
Manufacturer Narrative
|
H10: the device was not received for evaluation however a drawing of the set was provided to indicate the location of the crack.The crack was located on the access blood header of the filter.The reported condition was verified.The probable cause was identified to be due to a shock during transport or storage.A batch review was conducted and there were no deviations found related to this reported condition during the manufacture of this lot.Should additional relevant information become available, a supplemental report will be submitted.
|
|