H3, h6: the navio drill, part number 101209, s/n (b)(6) , used in treatment was returned for evaluation.A relationship between the reported event and the device was not confirmed.Nothing was identified visually that contributed to the reported problem.A functional evaluation was performed.A review of manufacturing and service records indicate the device met all specifications upon release into distribution.A complaint history review for similar reported/confirmed complaints has identified prior events.Although the reported problem was not confirmed, a factor that may have contributed to the reported symptom may have been associated with inadequate irrigation of the device.Although no further containment or corrective action is recommended or required at this time, all complaints are monitored and trended through post market surveillance activities.
|