• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MAKO SURGICAL CORP. HANDPIECE MICS ORTHOPEDIC STEREOTAXIC INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MAKO SURGICAL CORP. HANDPIECE MICS ORTHOPEDIC STEREOTAXIC INSTRUMENT Back to Search Results
Model Number 209063
Device Problems Failure to Calibrate (2440); Appropriate Term/Code Not Available (3191)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/20/2021
Event Type  malfunction  
Manufacturer Narrative
Serial specific voluntary recall was initiated for the mako integrated cutting system (mics) within scope of a capa. The initial root cause analysis determined that the process for characterizing mics handpieces for specific serial numbers deviated from its qualified state at the time of validation. The capa investigation is currently in progress and an updated communication will be submitted upon completion of the investigation.
 
Event Description
Handpiece would not verify during registration. Surgical delay:
=
15 minutes. Case type / application: tka.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameHANDPIECE MICS
Type of DeviceORTHOPEDIC STEREOTAXIC INSTRUMENT
Manufacturer (Section D)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer (Section G)
MAKO SURGICAL CORP.
2555 davie road
fort lauderdale FL 33317
Manufacturer Contact
daniela davila
2555 davie road
fort lauderdale, FL 33317
9546280700
MDR Report Key11992489
MDR Text Key255981675
Report Number3005985723-2021-00103
Device Sequence Number1
Product Code OLO
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Reporter Occupation
Remedial Action Recall
Type of Report Initial,Followup
Report Date 10/01/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/14/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Model Number209063
Device Catalogue Number209063
Device Lot Number420204201
Was Device Available for Evaluation? Yes
Is the Reporter a Health Professional? No
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received09/07/2021
Was Device Evaluated by Manufacturer? No
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Removal/Correction NumberZ-0472-2021

Patient Treatment Data
Date Received: 06/14/2021 Patient Sequence Number: 1
-
-