DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Unexpected Shutdown (4019)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 05/20/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A getinge field service engineer (fse) was dispatched to evaluate the iabp and was unable to reproduce the reported issue.However, the fse was able to verify the error codes of 111 and 112 which pointed towards replacing the executive processor board as required.To fix the issue, the fse replaced the executive board, and performed all functional and safety checks to meet factory specifications.Unit passed all functional and safety test per factory specifications.The iabp was then released to the customer and cleared for clinical service.A supplemental report will be submitted when additional information is provided.
|
|
Event Description
|
It was reported that during use on a patient, the cardiosave intra-aortic balloon pump (iabp) shut off and screen went blank.The unit was swapped with another to continue therapy.No patient harm, serious injury or adverse event was reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields: b4, b6, b7, g4, g7, h2, h6(investigation type, investigation findings & investigation conclusions), h10, h11.Corrected fields: g1(contact person).
|
|
Search Alerts/Recalls
|
|
|