Model Number 0684-00-0480-01
Device Problems Leak/Splash (1354); Off-Label Use (1494)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/25/2021
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.If provided we will send a supplemental report with our additional findings.Complaint record id # (b)(4).
 
Event Description
It was reported that during intra-aortic balloon (iab) therapy, the console generated a leak in iab circuit alarm.The alarm became more frequent going off approximately every five minutes.The insertion was reported to be axillary, which is not the method described in the device instructions for use.There was no patient harm or adverse event reported.
 
Manufacturer Narrative
The device has not been returned to the manufacturer so we are unable to complete an evaluation.Analysis of production (3331/213) - the device history records review concluded that there were no ncmrs, rework, or deviations documented for the reported lot/serial number.Based on the review results, it was determined that there is no relation between the manufacturing process and the reported failure.Historical data analysis (4109/213) - the review of the historical data was performed.Trend analysis (4110/213) - the overall complaint trend data for the period jun-19 to may-21 was reviewed.Communication/interviews: (4111/213) communication/interviews were performed to obtain all possible information.Reference complaint #: (b)(4).H3 other text : device not returned.
 
Event Description
N/a.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LINEAR 7.5 FR. 40CC IAB
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
Manufacturer (Section D)
DATASCOPE CORP. - FAIRFIELD
15 law drive
fairfield NJ
MDR Report Key11998039
MDR Text Key256182938
Report Number2248146-2021-00378
Device Sequence Number1
Product Code DSP
UDI-Device Identifier10607567106564
UDI-Public10607567106564
Combination Product (y/n)N
PMA/PMN Number
K041281
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 08/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/15/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/01/2023
Device Model Number0684-00-0480-01
Device Catalogue Number0684-00-0475
Device Lot Number3000129449
Was Device Available for Evaluation? Yes
Date Manufacturer Received08/18/2021
Patient Sequence Number1
Treatment
UNKNOWN.; UNKNOWN.
-
-