• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number UNKNOWN
Device Problem Microbial Contamination of Device (2303)
Patient Problem Bacterial Infection (1735)
Event Date 06/01/2019
Event Type  Injury  
Manufacturer Narrative
Patient identifier: patient information was not provided.Livanova (b)(4) implemented a field safety notice for disinfection and cleaning of heater-cooler devices.The z number is z-2076/2081-2015.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).Livanova initiated an investigation.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova received report that a male patient underwent cardiac surgery on (b)(6) 2019 and an heater-cooler system 3t was used.The serial number of device used during surgery is unknown.The patient showed signs of infection and illness.Reportedly, the patient was subjected to treatment and analysis of an infection in his chest.It was reported that his physicians reasonably concluded an ntm infection, but no ntm testing or results are mentioned.The patient continues to undergo treatment.
 
Event Description
See initial report.
 
Manufacturer Narrative
H.10: no dhr and shr could be performed since no serial number was provided.Through follow-up communication with the chief perfusionist under previous cases from the same hospital, livanova deutschland learned that the water in the heater-cooler systems 3t in use is changed every day and they are stored dry.This is not in alignment with current instruction for use however reportedly the devices in use at the hospital are very clean and there is no sign of biofilm.The devices are located inside the operating theater during use.The result of microbial sampling performed at customer site revealed that two devices in use at the hospital were found to be contaminated.It is not possible to determine if the device used for this specific surgery was one of the two devices confirmed to be contaminated and is not possible to retrieve the information if the device was already contaminated at the time of the surgery.No additional information has been provided regarding this specific case.The root cause of the event remains unknown.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstrasse 25
munich
MDR Report Key12006576
MDR Text Key257208867
Report Number9611109-2021-00317
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
PMA/PMN Number
K191402
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Remedial Action Recall
Type of Report Initial,Followup
Report Date 06/16/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberUNKNOWN
Was the Report Sent to FDA? No
Date Manufacturer Received09/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-2076/2081-2015
Patient Sequence Number1
Patient Outcome(s) Other;
-
-