ABBOTT VASCULAR G4 STEERABLE GUIDING CATHETER CATHETER, STEERABLE Back to Search Results
Model Number SGC0701
Device Problems Break (1069); Positioning Failure (1158)
Patient Problem Perforation (2001)
Event Date 05/27/2021
Event Type  Injury  
Manufacturer Narrative
The customer reported the device is not returning. Investigation is not yet complete. A follow-up report will be submitted with all additional relevant information.
 
Event Description
This is filed to report a perforation, unable to straighten the device and a cable break. It was reported that this was a mitraclip procedure to treat degenerative mitral regurgitation (mr) with a grade of 4+. The steerable guide catheter was inserted, but it was observed that when applying minus knob, the guide was unable to straighten. The physician suspected a cable break occurred; therefore, the clip was removed and replaced. It was noted that after the sgc was removed an atrial septal defect was observed. The physician decided no treatment was needed; therefore, a new sgc was inserted and one clip was successfully implanted, reducing mr to a grade of <1. There was no clinically significant delay in the procedure. No additional information was provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameG4 STEERABLE GUIDING CATHETER
Type of DeviceCATHETER, STEERABLE
Manufacturer (Section D)
ABBOTT VASCULAR
26531 ynez rd.
temecula CA 92591 4628
Manufacturer (Section G)
ABBOTT VASCULAR, REG # 3005070406
3885 bohannon drive
menlo park CA 94025
Manufacturer Contact
lindsey bell
26531 ynez rd.
temecula, CA 92591-4628
9519143996
MDR Report Key12009407
MDR Text Key256504338
Report Number2024168-2021-05037
Device Sequence Number1
Product Code DRA
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K190167
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 07/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date01/28/2022
Device Model NumberSGC0701
Device Catalogue NumberSGC0701
Device Lot Number10129U126
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received06/30/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/29/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 06/16/2021 Patient Sequence Number: 1
-
-