MAUDE Adverse Event Report: ARTHROCARE CORP. DISP FIRSTPASS STR PASSR SELF; PASSER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. DISP FIRSTPASS STR PASSR SELF; PASSER Back to Search Results
Catalog Number 22-4038
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/27/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event Description
It was reported that during an arthroscopy, a firstpass piece of the capture jaws broke in the patient.The piece was recovered.The procedure was completed with no significant delay using a back-up device.No other complications were reported.
 
Manufacturer Narrative
H10: h3, h6: the shipping information was provided; however, the subject device was not made available to the designated complaint unit and thus a physical product evaluation could not be performed.A device deficiency was not identified, and the root cause of the reported event could not be determined since the device was not returned for evaluation.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specifications upon release for distribution.A complaint history review found similar reported events.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure and/or harm was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the disposable firstpass assembly found that the manufacturer and the quality control inspector should perform a visual and functional final assembly inspection for upper and lower jaw.The root cause of the reported event could not be determined since the device was not received for evaluation.Please refer to the instructions for use for precautions and warnings related to the use of the device.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DISP FIRSTPASS STR PASSR SELF
Type of Device
PASSER
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer (Section G)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key12009661
MDR Text Key256531386
Report Number3006524618-2021-00715
Device Sequence Number1
Product Code HWQ
UDI-Device Identifier00885556724545
UDI-Public885556724545
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 01/22/2024
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/23/2024
Device Catalogue Number22-4038
Device Lot Number2066923
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received01/18/2024
Date Device Manufactured01/23/2021
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-