Model Number 8637-20
Device Problems Premature Elective Replacement Indicator (1483); Insufficient Flow or Under Infusion (2182); Inappropriate or Unexpected Reset (2959); Data Problem (3196)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/10/2021
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
Information was received from a consumer and a device manufacturer representative regarding a patient receiving bupivacaine (30 mg/ml at 7. 2 mg/day) and morphine (10 mg/ml at 2. 4 mg/day) via an implantable infusion pump. It was reported that the patient was getting alerts on his ptm while trying to deliver a bolus. He was at the office for a refill and per interrogation, 0ml should have been in the reservoir, but 5ml was aspirated. Codes 225 (safe state), 231 (pump reset), and 259 (pump memory error) were seen upon interrogation. The hcp was able to refill and update the pump successfully, however the pump continued to experience resets over the next 3 days. The pump reset and was reprogrammed 4 times. At the last reprogramming, the interrogation indicated that the estimated replacement interval was less than a month, despite the pump being implanted on (b)(6) 2021. They were able to program the pump to minimum rate using an 8840 programmer, then reprogram to the desired settings. The caller was advised to continue to monitor for alarms and alerts, and considerations were reviewed. The patient's weight was provided. No further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameSYNCHROMED II
Type of DevicePUMP, INFUSION, IMPLANTED, PROGRAMMABLE
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
road 31, km. 24, hm 4
ceiba norte industrial park
juncos PR 00777
Manufacturer Contact
glen belmer
7000 central avenue ne rcw215
minneapolis, MN 55432
6122713209
MDR Report Key12012546
MDR Text Key262474536
Report Number3004209178-2021-09513
Device Sequence Number1
Product Code LKK
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P860004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup,Followup,Followup,Followup
Report Date 10/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/16/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date04/28/2022
Device Model Number8637-20
Device Catalogue Number8637-20
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/15/2021
Is the Reporter a Health Professional?
Was the Report Sent to FDA?
Event Location No Information
Date Manufacturer Received10/12/2021
Was Device Evaluated by Manufacturer? No
Date Device Manufactured11/10/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial

Patient Treatment Data
Date Received: 06/16/2021 Patient Sequence Number: 1
-
-