Information was received from a consumer and a device manufacturer representative regarding a patient receiving bupivacaine (30 mg/ml at 7.
2 mg/day) and morphine (10 mg/ml at 2.
4 mg/day) via an implantable infusion pump.
It was reported that the patient was getting alerts on his ptm while trying to deliver a bolus.
He was at the office for a refill and per interrogation, 0ml should have been in the reservoir, but 5ml was aspirated.
Codes 225 (safe state), 231 (pump reset), and 259 (pump memory error) were seen upon interrogation.
The hcp was able to refill and update the pump successfully, however the pump continued to experience resets over the next 3 days.
The pump reset and was reprogrammed 4 times.
At the last reprogramming, the interrogation indicated that the estimated replacement interval was less than a month, despite the pump being implanted on (b)(6) 2021.
They were able to program the pump to minimum rate using an 8840 programmer, then reprogram to the desired settings.
The caller was advised to continue to monitor for alarms and alerts, and considerations were reviewed.
The patient's weight was provided.
No further complications were reported.
|