• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEROYAL INDUSTRIES, INC. DEROYAL; GENERAL SURGERY TRAY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEROYAL INDUSTRIES, INC. DEROYAL; GENERAL SURGERY TRAY Back to Search Results
Model Number 89-10571
Device Problem Contamination (1120)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 05/27/2021
Event Type  malfunction  
Event Description
During setup hair found inside big wrapped round blue basin inside the pack.Wrapped basin was discarded.The rest of the setup was considered not contaminated.No items saved.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
DEROYAL
Type of Device
GENERAL SURGERY TRAY
Manufacturer (Section D)
DEROYAL INDUSTRIES, INC.
200 debusk lane
powell TN 37849
MDR Report Key12016364
MDR Text Key256724499
Report Number12016364
Device Sequence Number1
Product Code LRO
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial
Report Date 05/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number89-10571
Device Catalogue Number89-10571
Device Lot Number53332666
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? Yes
Date Report Sent to FDA05/28/2021
Event Location Hospital
Date Report to Manufacturer06/17/2021
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-