ALCON RESEARCH, LLC - HOUSTON CONSTELLATION SURGICAL PROCEDURE PAK; GENERAL SURGERY TRAY (KIT) Back to Search Results
Catalog Number 8065751765
Device Problem Fluid/Blood Leak (1250)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/16/2021
Event Type  malfunction  
Manufacturer Narrative
Investigation including root cause analysis is in progress.A supplemental mdr will be filed as necessary in accordance with 21 cfr 803.56 when additional reportable information becomes available.The manufacturer internal reference number is: (b)(4).
 
Event Description
A nurse reported the cassette was noticed with water leakage before a vitrectomy procedure.The cassette was replaced and the procedure was completed.There was no patient harm.
 
Manufacturer Narrative
A review of the device history record traceable to the reported lot number indicates, that the product was processed and released according to the product¿s acceptance criteria.No sample has been returned for evaluation.Therefore, the condition of the product could not be verified.The root cause of the customer's complaint could not be established, as a sample has not been received.And the condition of the product could not be verified.Without analysis of the sample, it is not possible to isolate the root cause.As the root cause is unknown, the relationship if any, of the device to the reported incident cannot be determined.The root cause for this complaint is not known.Therefore, specific action with regards to this complaint cannot be taken.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Event Description
There was no patient involvement.
 
Manufacturer Narrative
(b)(4).The wet product was visually inspected and it was observed that one of the identification (id) tabs was broken off.A calibrated console representing the current software version was used to test the sample.A system message code displayed on the console screen to state: "invalid cassette id detected.Please eject and insert a valid cassette." the submerge leak test was performed on the cassette and no leakage was observed.The root cause of the customer's complaint could not be determined conclusively as no leakage was detected.The broken id tab issue would not have resulted in product leakage.Action will not be taken for this occurrence.After investigation of this complaint and analysis of complaints of this nature confirm no unfavorable trend for this event.Quality assurance has reviewed this complaint and will continue to monitor data for evidence of adverse trending and take further action, as appropriate.The manufacturer internal reference number is: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CONSTELLATION SURGICAL PROCEDURE PAK
Type of Device
GENERAL SURGERY TRAY (KIT)
Manufacturer (Section D)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer (Section G)
ALCON RESEARCH, LLC - HOUSTON
9965 buffalo speedway
houston TX 77054
Manufacturer Contact
jonathan schlech
6201 south freeway
mail stop ab2-6
fort worth, TX 76134
8175514979
MDR Report Key12017079
MDR Text Key256735949
Report Number1644019-2021-00405
Device Sequence Number1
Product Code LRO
UDI-Device Identifier00380657517657
UDI-Public00380657517657
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K880961
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Health Professional,Company Representative
Reporter Occupation Nurse
Type of Report Initial,Followup,Followup
Report Date 12/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date09/30/2022
Device Catalogue Number8065751765
Device Lot Number2423678H
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/06/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received11/12/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/29/2020
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
-
-