Catalog Number 05.001.080
Device Problems Output below Specifications (3004); Noise, Audible (3273); Complete Loss of Power (4015)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 01/01/2021
Event Type  malfunction  
Manufacturer Narrative
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was determined to be due to component failure.(b)(4).
 
Event Description
It was reported from (b)(6) that during service and evaluation, it was determined that the pen drive device had an air leak, excessive noise, worn motor and insufficient low power.It was further determined that the device failed pretest for check speed with the tachometer, internal leak tightness and external leak tightness.It was noted in the service order that the device did not work.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.The exact date of the event was unknown.However, it was reported that the event occurred in 2021.All available information has been disclosed.If additional information should become available, an additional report will be submitted accordingly.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AIR PEN DRIVE
Type of Device
INSTRUMENT, SURGICAL, ORTHOPEDIC, AC-POWERED, MOTOR/ACCESS AND ATTACHMENT
Manufacturer (Section D)
SYNTHES GMBH
eimattstrasse 3
oberdorf 4436
SZ  4436
MDR Report Key12024635
MDR Text Key280531436
Report Number8030965-2021-05062
Device Sequence Number1
Product Code HWE
UDI-Device Identifier07611819830899
UDI-Public07611819830899
Combination Product (y/n)N
PMA/PMN Number
K093361
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial
Report Date 04/06/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number05.001.080
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/06/2021
Date Manufacturer Received06/09/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-