Catalog Number 530.710 |
Device Problems
Energy Output Problem (1431); Output below Specifications (3004); Complete Loss of Power (4015); Excessive Heating (4030)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 01/01/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: as of this date, the device has not been returned for evaluation; therefore, the reported condition cannot be confirmed and/or duplicated.(b)(4).
|
|
Event Description
|
It was reported from (b)(6) that during an unspecified surgical procedure of the spine, it was observed that the battery oscillator device was heating up, powerless and could not be used cut bone.There were no delays in the surgical procedure as an identical spare device was available for use.There was patient involvement.There were no injuries, medical intervention or prolonged hospitalization.The date of event was unknown but was known to have occurred in 2021.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: service history review: the actual device was not returned for evaluation; therefore, the reported condition was not confirmed.However, a service history review was performed which indicated that the device was serviced within the last year for a service condition that is not relevant to the current reported condition.
|
|
Manufacturer Narrative
|
This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes, or its employees that the report constitutes an admission that the product, depuy synthes, or its employees caused or contributed to the potential event described in this report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.H10 additional narrative: the actual device was returned for evaluation.The device was evaluated and it was determined that the reported condition that the device was overheating was not confirmed.During service and evaluation it was observed that the device would not run.The device also failed pretests for function of device and oscillation frequency with frequency meter.The assignable root cause was traced to component failure due to normal wear.
|
|