OLYMPUS MEDICAL SYSTEMS CORP. ULTRASONIC PROBE Back to Search Results
Model Number UM-S20-17S
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device in this report has not been returned to olympus medical systems corp.(omsc) for evaluation.The exact cause of the reported event could not be conclusively determined at this time.If additional information becomes available, this report will be supplemented.
 
Event Description
At the end of the bronchoscopy, the user found that a small piece of the distal end of the ultrasound probe fell off in the patient's body.The user removed the small piece from the patient body.The user took an x-ray analysis for the patient and confirmed that the piece was not left in the patient's body.There was no extension of the procedure relevant clinically.There were no reports of patient injuries related to this incident.The user facility did not provide other detailed information.
 
Manufacturer Narrative
This supplemental report is being submitted to provide additional information.The device was evaluated at olympus repair center.According to the evaluation, olympus repair center found that the fractured part of the device was not flat.Device history record review indicates that the product was manufactured and tested in accordance with all applicable procedures and met all final product release criteria.The exact cause of the reported event could not be conclusively determined.Omsc surmised that this phenomenon attributed to the applied external stress to the device.If significant additional information is received, this report will be supplemented.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ULTRASONIC PROBE
Type of Device
ULTRASONIC PROBE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key12038313
MDR Text Key257458635
Report Number8010047-2021-07811
Device Sequence Number1
Product Code ITX
Combination Product (y/n)N
PMA/PMN Number
K982323
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup
Report Date 09/10/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberUM-S20-17S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/17/2021
Was the Report Sent to FDA? No
Date Manufacturer Received08/25/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-