Model Number 71365705
Device Problem Break (1069)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/07/2021
Event Type  malfunction  
Event Description
It was reported that, during the implantation of the stem in a total hip replacement, an anthology inserter posterior hard broke outside the patient.Surgery was resumed, without any delay, with a smith & nephew back-up device.Patient was not injured as consequence of this problem.
 
Manufacturer Narrative
The device, intended for use in treatment, was returned and evaluated.A visual inspection of the returned anthology inserter poster hard confirmed the top of the device is broken off.The device also shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ANTHOLOGY INSERTER POSTER HARD
Type of Device
PROSTH, HIP, SEMI-CONSTRA, UNCEMENT, METAL/POLY, NON-POROUS, CALICUM-PHOSPHATE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key12042998
MDR Text Key257488691
Report Number1020279-2021-05339
Device Sequence Number1
Product Code MEH
UDI-Device Identifier03596010555663
UDI-Public03596010555663
Combination Product (y/n)N
PMA/PMN Number
K113789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 07/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/22/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71365705
Device Catalogue Number71365705
Device Lot Number19GM09428
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/25/2021
Date Manufacturer Received06/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-