Model Number 201
Event Type  Malfunction  
Event Description

Reporter indicated a vns wand was not communicating despite changing the wand battery. The wand was returned for analysis. Product analysis noted that the serial data cable, which produced communication errors, had open and intermittent conductors. A known good bench serial data cable was substituted and all communications errors cleared.

 
Manufacturer Narrative

Device failure occurred, but did not cause or contribute to a death or serious injury.

 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NamePROGRAMMING WAND
Type of DevicePROGRAMMING WAND
Manufacturer (Section D)
CYBERONICS, INC.
houston TX 77058
Manufacturer Contact
nydia grimes
100 cyberonics blvd.
ste. 600
houston , TX 77058
2812287200
MDR Report Key1204676
Report Number1644487-2008-02530
Device Sequence Number1
Product CodeLYJ
Report Source Manufacturer
Source Type Health Professional,Company Representative
Reporter Occupation
Type of Report Initial
Report Date 09/27/2008
1 Device Was Involved in the Event
0 PatientS WERE Involved in the Event:
Date FDA Received10/20/2008
Is This An Adverse Event Report? No
Is This A Product Problem Report? Yes
Device Operator HEALTH PROFESSIONAL
Device MODEL Number201
Was Device Available For Evaluation? Yes
Date Returned to Manufacturer10/07/2008
Is The Reporter A Health Professional? Yes
Date Manufacturer Received09/27/2008
Was Device Evaluated By Manufacturer? Yes
Date Device Manufactured04/01/2005
Is The Device Single Use? No
Is this a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Invalid Data

-
-