Device evaluated by mfr: synergy ous mr 3.00mm x 38mm stent delivery system, catheter was returned for analysis.Examination of the stent identified the stent was fully separated from the delivery device.Stent damage and stretching was identified along the entire length of the stent.A review of the manufacturing stent profile data was performed and the stent outer diameter was within max crimped stent profile measurement.The balloon cones were reviewed, and the balloon appeared to have been subjected to positive pressure and the balloon cones were relaxed.There was clear dried media inside the balloon and there were no signs of damage to the balloon.Examination of the tip found no damage.Examination of the hypotube found no issues.A visual and tactile examination of the outer and mid-shaft section and a visual examination of the inner lumen found no issues.Device was tracked on 0.014 guidewire without issue.Device inflated with no issues to rated burst pressure, maintained pressure and deflated within 16 seconds.No other issues were identified during the product analysis.
|
Reportable based on device analysis completed on17jun2021.It was reported that stent is having difficulty in deployment.Lya 3.00 x 38 synergy ii des drug-eluting stent was advanced, however, during the first inflation to nominal pressure stent did not deploy correctly.The procedure was completed with a different device.There were no complications reported and the patient is stable.However, returned device analysis revealed stent damage.
|