• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: WILLIAM COOK EUROPE UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

WILLIAM COOK EUROPE UNKNOWN; DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR Back to Search Results
Catalog Number UNKNOWN
Device Problems Difficult to Remove (1528); Obstruction of Flow (2423); Structural Problem (2506); Device Tipped Over (2589); Unintended Movement (3026); Appropriate Term/Code Not Available (3191)
Patient Problems Dyspnea (1816); Internal Organ Perforation (1987); Pain (1994); Weight Changes (2607); Movement Disorder (4412); Thrombosis/Thrombus (4440); Insufficient Information (4580)
Event Type  Injury  
Manufacturer Narrative
Occupation: non-healthcare professional.Investigation: it has not been possible to further investigate or evaluate this alleged event based on the limited information and/or no device failure provided to date.Catalog number and lot number are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up medwatch report will be submitted if additional relevant information becomes available.
 
Event Description
It is alleged that the patient received a tulip inferior vena cava (ivc) filter on (b)(6) 2011, and the patient was injured without further explanation.Hospital and medical records have been requested, but not yet provided.
 
Manufacturer Narrative
Investigation: investigation is reopened due to additional information provided.The reported allegations have been further investigated based on the information provided to date.The following allegations have been investigated: organ perforation, dvt, difficult to retrieve, migration, tilt, involuntary movement, pain, dyspnea, inability to lose weight, requires a caregiver, physical limitations, schizoaffective disorder.Filter interacts with ivc wall, e.G.Penetration/perforation/embedment.This may be either symptomatic or asymptomatic.Potential causes may include improper deployment; and (or) excessive force or manipulations near an in-situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: trauma to adjacent structures, vascular trauma, vena cava perforation, vena cava penetration.Ivc occlusion/ thrombosis, new dvt, ivc stenosis as a reported complication, is a known risk in relation to filter implant and is well documented in the clinical literature and in clinical practice guidelines.This is supported by the clinical evidence report established to assess available clinical data to identify and evaluate the clinical safety and performance of the cook vena cava filters.Potential adverse events that may occur include, but are not limited to, the following: vena cava occlusion or thrombosis, vena cava stenosis, deep vein thrombosis.Physician practice guidelines and published guidance from regulatory agencies recommend that patients with indwelling filters undergo routine follow-up.The risks/benefits of filter retrieval should be considered for each patient during follow-up.Once protection from pe is no longer necessary, filter retrieval should be considered.Filter retrieval should be attempted when feasible and clinically indicated.Filter retrieval is a patient-specific, clinically complex decision; the decision to remove a filter should be based on each patient¿s individual risk/benefit profile (e.G., a patient¿s continued need for protection from pe compared to their experience with and (or) ongoing risk of experiencing filter-related complications).For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth.The filter is designed to be retrieved with the günther tulip vena cava filter retrieval set.It may also be retrieved with the cloversnare® vascular retriever.Cook has not performed testing to evaluate the safety or effectiveness of filter retrieval using other retrieval systems or techniques.The published clinical literature includes descriptions of alternative techniques for filter retrieval; use of these techniques varies according to physician experience, patient anatomy, and filter position.The safety or effectiveness of these alternative retrieval techniques has not been established.Specific for ¿embedded¿ a filter that is embedded in the wall of the ivc may be difficult to retrieve.For all retrievable ivc filters, retrieval becomes more challenging with time, and this is commonly due to encapsulation of the filter legs or hook (in a tilted filter) by tissue ingrowth.Filter or filter fragment migration and (or) embolization (e.G., movement to the heart or lungs) has been reported.Filter or filter fragment movement has occurred in both the cranial and caudal direction and may be either symptomatic or asymptomatic.Potential causes may include filter placement in ivcs with diameters smaller or larger than those specified in these instructions for use; improper deployment; deployment into thrombus; dislodgement due to large thrombus burdens; and (or) excessive force or manipulations near an in situ filter (e.G., a surgical or endovascular procedure in the vicinity of a filter).Potential adverse events that may occur include, but are not limited to, the following: filter migration, trauma to adjacent structures.Filter tilt has been reported.Potential causes may include filter placement in ivcs with diameters larger than those specified in these instructions for use; improper deployment; manipulations near an implanted filter (e.G., a surgical or endovascular procedure in the vicinity of a filter); and (or) a failed retrieval attempt.Excessive filter tilt may contribute to difficult or failed retrieval; vena cava wall penetration/perforation; and (or) result in loss of filter efficiency.Potential adverse events that may occur include, but are not limited to, the following: unacceptable filter tilt.Unknown if the reported involuntary movement, pain, dyspnea, inability to lose weight, requires a caregiver, physical limitations, schizoaffective disorder is directly related to the filter and unable to identify a corresponding failure mode at this point in time.Catalog and lot numbers are unknown.The alleged tulip is manufactured and inspected according to controls.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Patient allegedly received a filter on (b)(6) 2011 via the right internal jugular vein due to blood clots - vitamin k deficiency.Patient is alleging device migration and tilt, is unable to be retrieved and organ perforation.Patient notes and further alleges experiencing "involuntary movement, breathing issues, stomach pains, chest pains, leg pain, back pain, neck pain", post implant deep vein thrombosis (dvt), inability to lose weight, requires a caregiver, physical limitations, and schizoaffective disorder.
 
Manufacturer Narrative
Blank fields on this form indicate the information is unknown, unavailable, or unchanged.Investigation: investigation is reopened due to additional information provided.Per quality engineering review, the additional information provided for this complaint does not change the previous investigation conclusion.Therefore, no new investigation activities will be conducted at this time.Cook will reopen the investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.Catalog and lot number are unknown, however, the alleged tulip is manufactured and inspected according to specifications.No evidence to suggest that this device was not manufactured according to specifications and nothing indicates that the filter did not perform as intended, e.G.Intended for the prevention of recurrent pulmonary embolism (pe) via placement in the vena cava.Cook will reopen its investigation if further information is received warranting supplementation in accordance with 21 c.F.R.803.56.This report includes information known at this time.A follow-up report will be submitted should additional relevant information become available.This report is required by the fda under 21 cfr part 803 and is based on unconfirmed information submitted by others.Neither the submission of this report nor any statement contained herein is intended to be an admission that any cook device is defective or malfunctioned, that a death or serious injury occurred, nor that any cook device caused, contributed to, or is likely to cause or contribute to a death or serious injury if a malfunction occurred.
 
Event Description
Per computed tomography report, "cook gunther tulip filter with: 26 mm vertebral, 7 mm small bowel, and 4 mm mesenteric perforation.Tilting with the apex against the ivc wall.No fracture, stenosis, or migration.".
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNKNOWN
Type of Device
DTK FILTER, INTRAVASCULAR, CARDIOVASCULAR
Manufacturer (Section D)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer (Section G)
WILLIAM COOK EUROPE
sandet 6, dk-4632
bjaeverskov
Manufacturer Contact
lissi walmann
sandet 6, dk-4632
bjaeverskov 
56868686
MDR Report Key12055274
MDR Text Key259387942
Report Number3002808486-2021-01385
Device Sequence Number1
Product Code DTK
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Consumer
Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date 06/27/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/23/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberUNKNOWN
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? No
Date Manufacturer Received06/14/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Life Threatening;
Patient SexMale
Patient Weight172 KG
-
-