MAUDE Adverse Event Report: ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP CRT-P IMPLANTABLE PACEMAKER PULSE GENERATOR

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ST. JUDE MEDICAL, INC.(CRM-SYLMAR) QUADRA ALLURE MP CRT-P IMPLANTABLE PACEMAKER PULSE GENERATOR Back to Search Results
Model Number PM3562
Device Problems Failure to Capture (1081); High impedance (1291)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/11/2021
Event Type  malfunction  
Event Description
It was reported that the patient presented for a routine implant. It was observed during implant that the left ventricular (lv) pacing lead impedance was high and out of range and no capture was present on the pacemaker. The lv lead was unscrewed and attached to a pacing system analyzer and tested normal. The lead was then hooked back to the pacemaker and tested with different vectors. A few of the vectors produced in range impedance while others did not. The lead was then connected to a different pacemaker and tested normal. The first pacemaker was exchanged, the leads were connected to the new pacemaker and everything tested normal. The patient was stable.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameQUADRA ALLURE MP CRT-P
Type of DeviceIMPLANTABLE PACEMAKER PULSE GENERATOR
Manufacturer (Section D)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer (Section G)
ST. JUDE MEDICAL, INC.(CRM-SYLMAR)
15900 valley view court
sylmar CA 91342
Manufacturer Contact
sonali arangil
15900 valley view court
sylmar, CA 91342
MDR Report Key12063994
MDR Text Key258307041
Report Number2017865-2021-24056
Device Sequence Number1
Product Code NKE
UDI-Device Identifier05414734510097
UDI-Public05414734510097
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P030035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation
Type of Report Initial,Followup
Report Date 07/27/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/24/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Device Expiration Date04/30/2022
Device Model NumberPM3562
Device Catalogue NumberPM3562
Device Lot NumberP000111444
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/18/2021
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received07/13/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/21/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-