Model Number PM3562 |
Device Problems
Loose or Intermittent Connection (1371); Contamination of Device Ingredient or Reagent (2901)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/22/2021 |
Event Type
malfunction
|
Event Description
|
It was reported the patient presented in the hospital for a pacemaker implant procedure.It was observed there was a clicking sound coming off the torque wrench when connecting to the a-lead.The set screw was became stripped.It was also noted that silicone chips may have entered the screw holes.The physician elected to use another pacemaker to complete the procedure.The patient experienced no symptoms related to this procedure.
|
|
Event Description
|
It was reported the patient presented in the hospital for a pacemaker implant procedure.It was observed the set screw became stripped.It was also noted that silicone chips may have entered the screw holes.The physician elected to use another pacemaker to complete the procedure.The patient experienced no symptoms related to this procedure.
|
|
Manufacturer Narrative
|
Correction: h6.
|
|
Manufacturer Narrative
|
Customer complaint of stripped a was confirmed.Visual examination of the header found atrial septum and setscrew damaged and stripped.Left ventricular setscrew was removed and not returned, lv septum had setscrew particles consistent with stripping and removal of setscrew.Stripping of setscrews are consistent with overturning of the torque wrench and inserting the torque wrench at an angle.Performed dhr review to ensure that each manufacturing and inspection operation was performed.The product passed all quality control tests prior to its distribution.The pacer was received in normal working conditions with battery voltage above eri.Electrical analysis performed, indicated normal device functionality.Longevity assessment was performed and device was in the normal range of operation with appropriate remaining longevity.
|
|
Search Alerts/Recalls
|