Device Problem Gas Leak (2946)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 06/21/2021
Event Type  malfunction  
Manufacturer Narrative
No report of patient involvement. The initial reporter is located outside the u. S. , and therefore this information is not provided due to country privacy laws. A ge healthcare service representative performed a checkout of the system and confirmed the reported issue. The inspiratory flow sensor was replaced to resolve the issue.
 
Event Description
The hospital reported the inspiratory flow sensor detached in the middle resulting in a leak of 6l/min. There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand NameAESPIRE 7900
Type of DeviceANESTHESIA GAS MACHINE
Manufacturer (Section D)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer (Section G)
DATEX-OHMEDA, INC.
3030 ohmeda dr,
madison, WI 53718
Manufacturer Contact
john szalinski
3030 ohmeda drive
madison, WI 53718
MDR Report Key12071177
MDR Text Key261436902
Report Number2112667-2021-01625
Device Sequence Number1
Product Code BSZ
Combination Product (y/n)N
PMA/PMN Number
K172045
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional,u
Type of Report Initial
Report Date 06/25/2021
1 Device was Involved in the Event
0 Patients were Involved in the Event:
Date FDA Received06/25/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Event Location No Information
Date Manufacturer Received06/21/2021
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured09/06/2007
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage

-
-