The patient was on continuous renal replacement therapy.The cycler suddenly stopped and alarmed, per report of the rn operating the device.The alarm was unable to be muted and unable to stop/re-start.The plug was pulled to stop the alarming, then re-inserted the plug and attempted re-start, but alarm resumed.Unable to return blood to the patient and he lost 310 ml of blood and required a transfusion later in the day for low hematocrit.Three additional set-up were attempted with the same cartridge lot # but none allowed rn to program or prime.A different lot number cartridge was then attempted, and it was set up and running without issue.Lot # 01078004 was the number for the 1st 3 setups.Lot # 10378017 worked.
|