Device Problem Application Program Version or Upgrade Problem (2881)
Patient Problem Insufficient Information (4580)
Event Type  malfunction  
Event Description
It was reported that during a transport flight to a referring facility, the healthcare provider turned on the medfusion pump 4000 to prepare a fentanyl drip infusion.The medfusion pump screen stated the following: "confirm system update." the operator could not preform the updat as they were no longer within wifi proximity of the hospital.There was no option to bypass from the screen that was displayed.The operator turned off the pump and then restarted it to only have the medfusion pump display the same "confirm system update" message.This time the operator tried to do the update, but the screen continued to state, "updating system".From the "update system" screen the operator only had the following options: (1) update or (2) biomed.As the operator was still in flight to the referring facility, they could not reach the other medfusion pump 4000, so they decide to use the medfusion pump 3500 that was only reserved for epi infusions.When the operator later reached the facility, they were able to reach the secondary medfusion pump 4000, but it too required an update.Additional information was requested but has not yet been received.Should additional relevant details become available, a supplemental report will be submitted.
 
Manufacturer Narrative
Other, other text: this remediation mdr was generated under protocol b10010116, as a result of warning letter cms# 617147.Device evaluation this report addresses the event that the pump library update was delayed almost 48 hours.No product sample was received; therefore, visual, and functional testing could not be performed.A device history record (dhr) review could not be performed as the serial number was unknown.No root cause could be determined as the complaint could not be confirmed.Additional information h6., corrected data: corrections d1, d2, and g5.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MEDFUSION 4000 WIRELESS SYRINGE INFUSION PUMP
Type of Device
PUMP, INFUSION
Manufacturer (Section D)
ST PAUL
1265 grey fox rd.
st. paul MN 55112
Manufacturer (Section G)
NULL
1265 grey fox rd.
st. paul MN 55112
Manufacturer Contact
david halverson
6000 nathan lane north
minneapolis, MN 55442
MDR Report Key12078707
MDR Text Key261505067
Report Number3012307300-2021-06596
Device Sequence Number1
Product Code FRN
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
K111386
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type User Facility
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date 01/25/2023
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/28/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Date Manufacturer Received01/24/2023
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-