DATASCOPE CORP. - MAHWAH CARDIOSAVE HYBRID TYPE B PLUG; SYSTEM, BALLOON, INTRA-AORTIC AND CONTROL
|
Back to Search Results |
|
Model Number 0998-00-0800-53 |
Device Problem
Communication or Transmission Problem (2896)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 06/04/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
A supplemental report will be submitted when additional information is provided.
|
|
Event Description
|
It was reported by the flight nurse that an internal communication error alarm was displayed on the cardiosave intra-aortic balloon pump (iabp).Prior to call he had already attempted to reboot pump twice which did not work.As a result, he successfully switched consoles.The rn stated he would take pump out of service and mark for biomed.This pump was transferred and does not belong to this particular hospital.The rn stated he would handle the communication for surveillance.Limited information available.It is unknown the circumstances in which the event occurred or if there was patient involvement.No adverse event was reported.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields: b4, b6, b7, g4, g7, h2, h6 (type of investigation, component codes), h10, h11 corrected field: d4 (version or model#) testing of actual/suspected device (10) a getinge field service engineer (fse) was dispatched to evaluate the iabp.The fse i solated intermittent failure to coiled cable.To address the issue the fse replaced the coiled cable and replaced the expired battery modules.All functional and safety tests were passed to meet factory specifications, and the iabp was returned to the customer and cleared for clinical service.
|
|
Event Description
|
N/a.
|
|
Manufacturer Narrative
|
Updated fields: b4, g4, g7, g8, h2, h6 (type of investigation, investigation findings, investigation conclusions), h10, h11.Corrected field: g1 (contact person ¿ mfg site), h4.
|
|