• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE STRYDE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

NUVASIVE SPECIALIZED ORTHOPEDICS, INC. PRECICE STRYDE SYSTEM; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number PS10.0-50B235
Device Problem Corroded (1131)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned for evaluation.The root cause is unable to be determined at this time.If any additional information is provided, a supplemental report will be submitted.
 
Event Description
Information was received that after being explanted, the nail was noted to have corrosion on the proximal portion of the lengthening rod.No patient impact or adverse event was reported.No additional information is available.
 
Manufacturer Narrative
The returned nail was received for visual and functional testing.Visual inspection of the nail revealed that the distraction rod of the nail had visible evidence of discoloration at the junction.Per reported failure functional testing was not applicable.The root cause of the corrosion identified on the nail is currently being investigated.Review of the device history records indicated the precice-stryde nail was visually inspected, and functionally tested prior to release.
 
Event Description
No additional information has been provided.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PRECICE STRYDE SYSTEM
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
NUVASIVE SPECIALIZED ORTHOPEDICS, INC.
101 enterprise dr, suite 100
aliso viejo CA 92656
MDR Report Key12089621
MDR Text Key259629866
Report Number3006179046-2021-00322
Device Sequence Number1
Product Code HSB
UDI-Device Identifier00887517950741
UDI-Public887517950741
Combination Product (y/n)N
PMA/PMN Number
K180503
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received06/29/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPS10.0-50B235
Device Lot Number9092617
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer07/12/2021
Was the Report Sent to FDA? No
Date Manufacturer Received08/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-