The reported event could be confirmed, since the product was returned for evaluation and matches the alleged failure mode.The device inspection revealed the following: the device received shows traces of intense and frequent usage.The peg at the tip of the driver is completely broken off.The broken peg is not available for evaluation.The remnant peg shows typical signs of a forced, ductile breakage due to overload with evident material displacement.This indicates intense usage of the lag screwdriver with high torque application.It was manufactured in 2009, so it can be said that it had performed its function in the previous surgeries as intended without any reported failure.Since this has been long in use, thus a normal weakened structural integrity due to repeated reprocessing cycles is considered possible.A review of the device history for the reported lot did not indicate any abnormalities.No corrective actions are required at this time.No indications of material, manufacturing or design related problems were found during the investigation.A review of the labeling did not indicate any abnormalities.The instructions for use instructs user that: ¿always treat the instrument carefully to avoid surface damage or alterations to the geometry.The stryker "instructions for cleaning, sterilization, inspection and maintenance" help to determine whether an instrument is in good condition or must be replaced due to excessive wear or obvious defects.¿ based on investigation, the root cause was attributed to a normal wear related issue.The u-blade lag screwdriver was manufactured in 2009, so it can be said that it has performed its function in the previous surgeries as intended without any reported failure.Since it has been long in use, thus a normal weakened structural integrity due to repeated reprocessing cycles is considered possible.If any further information is provided, the complaint report will be updated.
|