MAUDE Adverse Event Report: ARTHROCARE CORP. 1.8MM Q-FIX ALL SUTURE ANCHOR; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. 1.8MM Q-FIX ALL SUTURE ANCHOR; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Catalog Number 25-1800
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Hypersensitivity/Allergic reaction (1907)
Event Date 06/09/2021
Event Type  Injury  
Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event Description
It was reported that after a case of humerus fracture or artificial joint surgery, an allergic reaction occurred from a patient who used q-fix all suture anchor.It is unknown how this was treated and the status of the patient.
 
Manufacturer Narrative
H10: h3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A clinical review states the information provided is insufficient to determine whether the patient¿s symptoms, signs or outcome are due to a pre-existing or concurrent medical or surgical condition or procedure, or to an adverse reaction to or experience with the device, one or more of its components, or its intended therapeutic action.Per the complaint, it is unknown how the allergic reaction was treated, and the status of the patient is unknown.Therefore, the causal relationship between the s&n device and the reported allergic reaction cannot be confirmed.No further clinical/medical assessment is warranted at this time based on information provided.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
1.8MM Q-FIX ALL SUTURE ANCHOR
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key12099584
MDR Text Key259378991
Report Number3006524618-2021-00731
Device Sequence Number1
Product Code MBI
UDI-Device Identifier00885556724606
UDI-Public885556724606
Combination Product (y/n)N
PMA/PMN Number
K133727
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 08/12/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received07/01/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue Number25-1800
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Date Manufacturer Received08/11/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-