The device was returned to zoll medical corporation.The customer's report was not replicated or confirmed.The device was put through extensive testing without duplicating the report.The device was recertified and returned to the customer.Reviewing the device logs, it is likely that case id (b)(4) is the event for the case.However, the case has no real information; there is no ecg signal, no electrode id, no errors, or evidence of trouble shooting.If the cpr adaptor is not applied (as seen by the onsite zoll representative), there would be no activity in the file as seen.It cannot firmly be determined that this was the cause of the report.There is not enough evidence to confirm or deny the report.The multifunction cable and accessories used were not returned as part of this investigation.Analysis of reports of this type has not identified an increase in trend.
|