Model Number V273 |
Device Problem
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
Patient Problem
Discomfort (2330)
|
Event Date 06/03/2021 |
Event Type
Injury
|
Manufacturer Narrative
|
This device has not been returned at this time.If it is returned, analysis will be performed, and this report will be updated.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) was found to be operating in safety mode.The patient was experiencing symptoms due to being in vvi unipolar pacing from safety mode.The device was subsequently explanted and replaced.No additional adverse patient effects were reported.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.
|
|
Event Description
|
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) was found to be operating in safety mode.The patient was experiencing symptoms due to being in vvi unipolar pacing from safety mode.The device was subsequently explanted and replaced.No additional adverse patient effects were reported.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.
|
|
Manufacturer Narrative
|
This device has not been returned at this time.If it is returned, analysis will be performed, and this report will be updated.Upon receipt at our post market quality assurance laboratory, this device was thoroughly inspected and analyzed.Interrogation of the device confirmed it was operating in safety mode due to system resets.It was confirmed that brady therapy remained available.The system resets occurred during a telemetry session and caused the device to enter safety mode.Engineers determined that this device was demonstrating behavior consistent with a high internal cell impedance within the battery.The high internal impedance resulted in the system resets due to temporary high-power consumption related to telemetry attempts and subsequent reversion to safety mode operation.Boston scientific has issued a field safety notice regarding ingenio extended life (el) and cardiac resynchronization therapy pacemaker (crt-p) devices that may exhibit this behavior.This particular device is included in the high impedance in ingenio extended life (el) pacemaker and crt-p advisory population.
|
|
Search Alerts/Recalls
|