Investigation summary: the complaint device was received at the service center and evaluated.
There was no allegation of malfunction against the device from the customer, defects were identified during service evaluation.
Per service reports, this complaint can be confirmed.
During the service evaluation the following defects were identified: the motor sticks and is rusty.
The cable resistance is out of tolerance.
Blade doesn't lock into shaver.
The motor cable and motor were replaced to resolve the issues.
After repair, the device was found to be working according to the specifications.
The faulty parts was identified as the root cause for the device failure during the service evaluation.
Manufacturing record evaluation is not required as the reported event is not associated with the manufacturing process and/or the potential cause of the defect cannot be associated to manufacturing.
At this point in time, no corrective action is required, and no further action is warranted.
Depuy mitek will continue to track any related complaints within this device family as a means of monitoring the extent with which this complaint is observed in the field.
|
It was reported by the affiliate in (b)(6) that during service evaluation, it was determined that the handpiece device failed the electrical safety test.
During in-house engineering evaluation, it was determined that the motor on the device was corroded as it was sticking.
There was no procedure nor patient involvement reported.
No additional information was provided.
|